- High Sensitive Model
- Frequency Selector Switch to eliminate the effect of harmonics.
- Three AC current ranges:4mA/40mA/100A.
- Resolution 0.001mA.
AC A (50/60Hz) | 4/40mA/100A ±1%rdg±5dgt (4/40mA) ±1%rdg±5dgt (0 – 80A) ±5%rdg (80.1 – 100A) |
---|---|
AC A (WIDE) | 4/40mA/100A ±1%rdg±5dgt [50/60Hz], ±2.5%rdg±10dgt [20Hz – 1kHz] (4/40mA) ±1%rdg±5dgt [50/60Hz], ±2.5%rdg±10dgt [40Hz – 1kHz] (0 – 80A) ±5%rdg [50/60Hz], ±10%rdg [40Hz – 1kHz] (80.1 – 100A) |
Maximum circuit voltage | 600V AC/DC (between line/neutral) 300V AC/DC (against earth) |
Conductor size | Ø40mm max. |
Frequency response | 20Hz – 1kHz (40Hz – 1kHz:100A) |
Effect of External Stray Magnetic Field | 2mA AC approx. in proximity to a 15mm-dia conductor carrying 100A AC |
Safety Standard | IEC 61010-1 CAT III 300V Pollution Degree 2 IEC 61010-2-032 |
Response Time | Approx. 2 seconds |
Power source | Two R03 or equivalent (DC1.5V) batteries |
Dimensions | 185(L) × 81(W) × 32(D)mm |
Weight | 290g approx. |
Included Accessories | 9097 (Carrying Case) R03 (1.5V) × 2 Instruction Manual |