PH Meter Wincom

Share to:

Wincom Company Ltd. manufactured digital pH meters are categorized by design type, technical specifications, and laboratory functionality. They specialize in water analysis, agricultural testing, and cosmetics quality control. [1, 2, 3, 4, 5]
Comparison of Popular Models

Model TypepH RangeResolution & AccuracyKey Features
Wincom PHS-25Benchtop / Tabletop0.00 to 14.00 pH0.01 pH pm0.05 pHMeasures pH, mV (ORP), and Temp (-9.9°C to 99.9°C); features composite E-201-C electrode.
Wincom PH-8414Portable0.00 to 14.00 pH0.01 pH pm 0.01 pHManual/automatic temperature compensation (0°C to 100°C); ideal for outdoor or field laboratory testing.
Wincom PHS-2FIntelligent Benchtop0.00 to 14.00 pH0.01 pH pm 0.01 pHFeatures computer control options, Automatic Temperature Compensation (ATC), and high-precision multi-parameter reading.
Wincom PH98108Pen / Pocket Type0.0 to 14.00 pH0.1 pH pm 0.05 pHUltra-portable waterproof body, battery-operated 4 x 1.5V, budget-friendly for field soil and water checks.


Step-by-Step Operation Guide (Standard PHS Benchtop Series)
To operate a standard benchtop Wincom pH meter accurately, follow these steps: 
  1. Power Initialization: Connect the instrument to its standard power source and switch the device ON to allow the electronic components to stabilize.
  2. Electrode Cleaning: Remove the composite glass electrode from its storage solution and rinse the tip thoroughly with deionized or distilled water. Blot dry gently with a lint-free tissue.
  3. Buffer Calibration: Immerse the electrode in standard calibration buffer solutions (typically pH 4.00, 7.00, or 10.00). Adjust the calibration knobs or execute the digital menu auto-calibration until the display accurately locks onto the buffer value.
  4. Sample Testing: Rinse the electrode again, submerge it into your liquid sample, and wait roughly 1 to 2 minutes until the numeric digital reading completely stabilizes. Record the final pH value alongside the measured temperature.
  5. Electrode Storage: Clean the tip post-measurement. Never store the probe in distilled water. Always submerge it back inside a protective cap filled with a 3 mol/L KCL storage solution to maintain the reference membrane functionality.

Reviews

There are no reviews yet.

Be the first to review “PH Meter Wincom”

Your email address will not be published. Required fields are marked *

Related Items

Scroll to Top

Leave Your Message