Features:
1. Throughput 60T/H.
2. 14 inch touch screen.
3. Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting.
4. 3D holographic scattergram displays the accurate 5 part differentiation of WBC.
5. Large storage capacity: 100,000 results(including histogram, scattergram, patient information).
Models:
| Model | BK-6310 |
| Throughput | 60 Tests/hour |
| Assay Items | 5 parts, 29 parameters, 3 histograms, 3 Dscattergram |
| Principle | Tri-angle laser scatter, Flow cytometrymethod, 3D scattergram analysis, Impedancemethodfor RBC and PLT counting, Cyanide-free method for HGB test |
| Test Mode | CBC mode, CBC+DIFF mode Venous whole blood, Capillary whole blood and Prediluted |
| Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT,PDW, P-LCR, P-LCC, NEU%, LYM%, MON%, EOS%,BAS%,NEU#, LYM#, MON#, EOS#, BAS# 4 Research parameter: ALY%, ALY#, IG%, IG# |
| Performance | Item | Linearity range | Carry Over | CV |
| WBC | (0~300.0×10^9)/L | ≤0.5% | ≤2.0% |
| RBC | (0~8.00×10^12)/L | ≤0.5% | ≤1.5% |
| HGB | (0~250)g/L | ≤0.5% | ≤1.5% |
| PLT | (0~3000×10^9)/L | ≤1.0% | ≤4.0% |
| Sample Volume | CBC+DIFF mode≤20μl CBC mode≤10μl |
| Storage | 10,000 results including histograms, scattergrams and patient information |
| Interface | 4 USB ports, 1LAN port Bi-direction LIS, support HL7 protocol, Internal RFID reader |
| Power Supply | AC220V±10% 50/60Hz; 110V±10% 60Hz; 400VA |
| Package Size | 670*590*790mm(W*D*H) |
| Gross Weight | 53kg |
| Reagent Package Size | 330*330*330mm , 400*290*240mm , 400*290*240mm |
| Reagent Gross Weight | 22KG 2KG, 1KG |