YCM’s 5th-generation GC is starting to be a new powerful leader in a GC market. With strongly enhanced APCs (Advanced Pneumatic Control), installation of three units of inlet and detector for each enables to extend the various GC configurations to perform complex analyses.
**YL6500 GC provides substantially reliable and precise data by every single part from the stable inlets to the extremely sensitive detectors as well as the innovative haptic touch pad strengthening user interface with a finger-tip.
**Moreover, an intuitive chromatography data system with user-friendly interface improves the ease of instrument control and a network (LAN) communication speeds up a data process.**
Features:
View-
– Enlarged haptic color LCD (5.7″”) showing all factors at a glance |
– Classic and lab-suitable design |
– Intuitive and user-friendly view by YL chromatography data system |
Verification-
– High speed data process by network (LAN) communication |
– Enhanced APCs for all modules |
– Upgraded column oven |
: Programming ramp/plateaus : 25/26 |
: Temperature Stability : ± 0.01 ℃ |
Value-
– Saving user’s valuable time with a high throughput autosampler |
– Improved column conditioning function |
: Automatic set of split flow up to 5 ml/min on column conditioning |
– Sensitive leak detection (H2) |
– Prevention of oven malfunction (Over heating) |
– Automatic stop when an oven door gets opened during operation |
Variety-
– Maximum no.of inlet / detector installation : 3 |
– Various inlets available : Capillary(Split/Splitless) / Packed / On-column inlet |
– Various detectors with high sensitivity available |
1) Flame ionization detector (FID): MDL < 1.5 pg C/sec |
2) Thermal conductivity detector (TCD): MDL < 2.5 ng /ml |
3) Micro-thermal conductivity detector (µTCD): MDL < 400 pg/ml |
4) Nitrogen phosphorus detector (NPD):MDL < 0.2 pg N/sec, <0.02 pg P/sec |
5) Flame photometric detector (FPD) : 1.0 pg S/sec, 0.2 pg P/sec |
6) Electron capture detector (ECD) : MDL < 10 fg/sec |
7) Pulsed discharge detector (PDD) |
8) Pulsed flame photometric detector (PFPD) |
9) Mass Spectrometer(MS) |
Specification:
Column Oven | ||
YL6500 GC Oven System Module | – Usable volume : 14 L – Cooling down option : – 80 °C ~ 450 °C (with LN2 cryogenic cooling) °C ~ 450 °C (with LCO2 cryogenic cooling) – Temperature set-point : 0.1 °C – Temperature programming : 25 ramps/26 plateaus – Maximum heating rate : 120 °C/min – Maximum run time : 9,999 min – Temperature program method : Maximum up to 20 – Temperature stability : ±0.01 °C (Isothermal), ±0.1 °C (Gradient) | |
Inlet | ||
Packed Inlet | – Maximum Temperature : 450 °C – Total flow setting range : 0.01~ 100 ml/min | – Pressure setting range : 0.001 ~ 100 psi – Pressure stability < ± 0.005 psi – Temperature stability < ±0.1 °C – Flow stability < ±0.05 ml/min – Temperature setpoint : 0.1 °C |
Capillary Inlet (Split/Splitless Inlet) | – Maximum Temperature : 450 °C – Total flow setting range : 0.01 ~ 400 ml/min N2 0.01 ~ 1000 ml/min He/H2 – Splitless time setpoint : 0.01 min | |
On-Column Inlet | – Maximum Temperature : 450 °C – Temperature programming up to 5 steps |
Detector (Data Acquisition Rate : 200 Hz) | |||
Maximum Temp. | MDL | Others | |
Flame Ionization Detector | 450 °C | 1.5 pg carbon/sec | Linearity : 107 |
Thermal Conductivity Detector | 400 °C | 2.5 ng/ml (Standard) 400 pg/ml (uTCD) | – Flow through cell : 4 Rhenium-Tungsten filaments – Filament protection |
Electron Capture Detector | 400 °C | 10 fg/sec | Linearity : > 104 |
Nitrogen Phosphorous Detector | 400 °C | < 0.2 pg N/sec (Azobenzene) < 0.02 pg P/sec (Parathion methyl) | Linearity for N : > 10 Linearity for P : > 1044 |
Flame Photometric Detector | 300 °C | < 1.0 pg S/sec < 0.2 pg P/sec | Linearity for S : Calibration curve is compulsory Linearity for P: > 105 |
Pulsed Discharge Detector | 400 °C | (PDHID) : – Organic compound : low ppb – Permanent gas : low ppm (PDECD) : 10-15 | (PDHID) : Linearity 105 (PDECD) : Linearity 10 |
Autosampler: | |
Liqid Sample Autosampler | ChroZen PAL RSI/RTC System All the Sampling Techniques Available in ONE Unit |
• Syringe Volume: 0.5, 1, 5, 10, 25, 50 and 100 µl • Tray Capacity – YL3000A : 121 vials, 2 ml – YL3100A : 15 vials, 2 ml – YL3200A : 209 vilas, 2 ml (2 removable racks) – ChroZen PAL LSI : 648 vials, 2 ml • Electrical Control : LAN and TTL, RS232 (Option) • Filling/Injection Speed : 1~100 µl/sec • Internal Standard Volume : as low as step of 0.1 µl • Washing Solvent Capacity : 6 x 10 ml vials | • Liquid Operation – Tray capacity: 648 vials, 2 ml – Repeatability: < 0.60 % RSD – Carry Over: < 0.004 % • Headspace Operation – Tray Capacity: 180 Vials (10/20ml) – Syringe Temperature: 40 – 150 °C – Carry Over: < 0.05 % – Repeatability: < 1.00 % RSD • SPME (Arrow) – Trace level sensitivity – High mechanical robustness |
YL2000H/YL2100H Headspace Autosampler | YL2800T/YL2850T all in All in One Unit |
• Tray Capacity – YL2000H : 42 Vials (20 ml) (Optional :6 and 10 ml) – YL2100H : 14 Vials (20 ml) (Optional : 10 ml) • Syringe Sizes: 2.5 (standard) ; optional ; 1 and 5 ml • Sampling Repeats: Up to 15 • Sampling Volume: Steps of 0.01 ml • Injection Speed: 0.5 – 100 ml/min | • Liquid Operation – Tray capacity: 121 vials, 2 ml • Headspace Operation – Tray Capacity: 42 vials (20ml); optional: 6 and 10 ml – Syringe Temperature: 40 – 150 °C • SPME (Fiber) as option -Fiber cleaning station (Temperature: 210-300 ℃) |
Versa Static Headspace Autosampler | HT3 Static & Dynamic Headspace Autosampler |
• Autosampler Capacity: 20-Position • Platen Heater : Single position; ambient to 200 °C • Vial Size : 22 mL vials • Sample Loop : (Standard) 1 mL Silco® coated loop. (Option) 100, 250, 500 μL as well as 2, 3, and 5 mL | • Autosampler Capacity : 60-position • Platen Heater : Up to 10 vials simultaneously heated up to 300 °C • Vial Size: 9, 12, 22 mL vials accepted |
Reviews
There are no reviews yet.